Product Parameters
Product Details
Product Introduction:
* The TU-1030 model successfully meets the strict requirements for high-precision and high-reliability measurements, suitable for various applications. It can be used in biological research, bioindustry, pharmaceutical analysis, pharmaceuticals, teaching and research, environmental protection, food hygiene, clinical testing, public health, and epidemic prevention.
* The highly automated instrument has automatic diagnostic and calibration recovery functions, ensuring stable operation and test data reliability.
* Core components are all imported parts, all optical lens surfaces are coated with SiO2 protective film, and the fully sealed monochromator structure makes instrument analysis more accurate, with enhanced stability and longer service life.
* The independently designed socket-type tungsten lamp eliminates the need for optical adjustment when changing the lamp, making instrument setup and maintenance simpler and more convenient.
Instrument Features:
* Chinese mainframe operating system with menu display, can directly show test data, easy, convenient, and fast to use.
* Data storage function: large memory capacity meets the storage requirements for daily standard curves and experimental data. By using USB ports to connect software-controlled printers, data and spectra can be printed directly, and test reports can be edited and printed.
* Optional accessories according to user requirements: single-slot holder, manual four-slot holder, automatic eight-slot holder, glass holder, 1cm cuvette holder, 5cm cuvette holder, etc.
* Measurement flexibility: spectral scanning, time scanning, and other flexible measurement requirements can be achieved through PC software control.
Technical Specifications and Basic Parameters:
* Wavelength range: 190–1100nm
* Spectral bandwidth: 2nm
* Wavelength accuracy: ±0.5nm
* Wavelength reproducibility: ≤0.2nm
* Transmittance accuracy: ±0.5% τ
* Transmittance repeatability: 0.2% τ
* Stray light: ≤0.05% τ (220nm, 360nm)
* Stability: 0.001A/h (500nm after warm-up)
* Photometric modes: transmittance, absorbance, concentration, energy
* Wavelength adjustment: automatic
* Photometric range: -0.3–3A
* Display: 128*64 LCD
* Detector: imported silicon photodiode
* Light source: imported tungsten lamp
* Power supply: AC 220V/50Hz, 110V/60Hz
* Power consumption: 120W
* Instrument dimensions: 520×400×220mm
* Main unit net weight: 16Kg
About Us
Company Introduction Corporate Culture Customer CaseNews
News Information Technical Article Video CenterService
Sales Network After-Sales ServiceNavigation
Products Solution Contact Us
010-89482732
Official Account
Mobile Browsing
Copyright @ 2026 Beijing Haotianhui Instrument Co., Ltd ICP: Beijing ICP No. 17002344-3
sitemap.xml Manage Login Technical Support: Chemical Instrument Network